Retiport
Visual Evoced Cortical
Potentials(VECP)
 
Scotipic ERG
 
Photopic ERG
 
Pattern ERG
 
 
EOG
DESCRIPTION
Retiport is a low-price PC-based, 2-4-8-channel system for measuring bioelectrical signals generated both from the retina and the optic nerve. The basic set comprises: computer device, operator monitor, printer, keyboard, mouse, bioelectrical amplifier, electrodes and checkerboard monitro. All electrical connections are easily to install.

The recorded curves are displayed on a operator monitor. By settingmarkers the system allows conputing latencies and amplitudes automatically. Data may be printed out or transmitted by an interface to another PC. The plain basic equipment already provides-in compliance with the ISCEL (International Society for Clinical Electrophysiolgy and Vision)-simple and rapid determination of VER, EOG, ERG and pattern ERG without strain for the patient.
 
Retiport
allows the following examinations:
Basic version
Electro-retinography(ERG)
(scotopic/photopic)
30Hz Flicker Response
Pattern ERG
(detection of early Glaucoma)
Visual Evoced Cortical Potentials (VECP)
 
Optional
Visial Acuity (sequential, according to Hajek und Zrenner)
Oculography(EOG) tast and slow
Nystagmography
Saccaden analyse
 
FUNCTIONS
VER Stimulator
- 15" VGA Monitor
- Checkerboard:4x4 - 128x128 squares in powers of to
- Pattern position:
full, half or quarter screen
- Pattern reversal
- Appearance / Disappearance
- Screen luminance : 80 cd/§³
- Contrast adjustable under computer control from 1 to 99%
 
 
 
Ganzfeld Stimulator Q 400
- ISCEV Standard
- full field globe 400 mm
- all functions full computer controlled
- with LED-Flash:
Standard intensity = 3.5 cds / §³
- intensity steps:-30 dB to 0 dB in steps of 5 dB
- Background intensity:30 cd / §³ ,
250 cd / §³ , 500 cd / §³
 
- Central Fixation LED and EOG LED
+/-30 grd-brightness computer control
 
Optional
- High Flash Xenon Strobe from
3.5 cds / §³ to 100 cds / §³
- Color filter for both:flash and background light
 
 
TECHNICAL DATA
¡á Power
- 115/230 V AC 50/60 Hz 450 VA
 
¡á Computer/Monitor/Printer :
- IBM compatible Pentium lll, 450 MHz
- Windows 98 or NT
- Monitor : 15"/17" operator monitor
- Printer : HP Desk Jet
 
¡á Amplifier / ADU / Interface :
- Common Mode Rejection > 110dB
- Noise < 4uV(RMS)
- Gain settings : steps 10uV - 10mV/DIV
- Low Cut Filter : 0.02 - 0.05 - 0.1 - 0.2 - 0.5 - 1 - 2 - 5 - 10 - 20 - 50 - 100 - 200 - 500 - 1000Hz
- High Cut Filter : 30 - 50 - 100 - 200 - 300 - 500 - 1K - 2K - 3K - 5K - 10K Hz
- Time Base:steps 50ms to 2000ms
- Trigger intern / extern
- 12 Bit ADU
- 2/4/8 channel System computer controlled